USE OF SOLID SAMPLING ZEEMAN ATOMIC-ABSORPTION SPECTROMETRY (SS-ZAAS)FOR CERTIFICATION PURPOSES

Citation
C. Hofmann et al., USE OF SOLID SAMPLING ZEEMAN ATOMIC-ABSORPTION SPECTROMETRY (SS-ZAAS)FOR CERTIFICATION PURPOSES, Fresenius' journal of analytical chemistry, 349(12), 1994, pp. 779-783
Citations number
23
Categorie Soggetti
Chemistry Analytical
ISSN journal
09370633
Volume
349
Issue
12
Year of publication
1994
Pages
779 - 783
Database
ISI
SICI code
0937-0633(1994)349:12<779:UOSSZA>2.0.ZU;2-S
Abstract
Three methods for the calibration of solid sampling Zeeman atomic abso rption spectrometry allowing independent calibration based on gravimet ric principles and thus usable for certification purposes are describe d and intercompared: standard addition to the solid sample, use of syn thetic solid matrix matching samples and extrapolation to zero matrix. lead to the best results, extrapolation to zero matrix is less suited at high concentrations, but is useful where an accuracy of the order of 10% is acceptable.