Study objective: We investigated the hydrodynamic characteristics of I
V infusion sets for rapid fluid resuscitation. A simple technique has
been devised for quantitative evaluation of the hydrodynamic character
istics of IV sets, including their components, for a range of infusion
pressures. Setting and methods: Previous investigations have measured
the overall flow rate of infusion sets with and without IV catheters.
This study presents a quantitative technique for measuring the resist
ance to flow of the IV delivery set as a whole as well as its componen
ts. An infusion set was measured with 14- and 18-gauge IV catheters wh
ile delivering fluid at infusion pressures between 50 (gravity) and 40
0 mm Hg. Measurements and main results: At gravity-driven infusion, th
e drip chamber imposes a resistance to flow of the same order as that
of the catheter. At pressurized infusion with small-bore catheters, th
e catheter consumes the majority of the overall pressure drop. At pres
surized infusion with a large-bore catheter or tubing, the standard dr
ip chamber becomes the limiting component and imposes the largest resi
stance to flow. Conclusion: At gravity-delivered pressures (50 and 100
mm Hg), the only effective way of increasing flow rate (more than two
fold) is to use a low-resistance drip chamber or to use two infusion s
ites. At pressurized delivery pressures (more than 200 mm Hg), increas
ing catheter size from 18 to 14 gauge would be more effective than dou
bling the number of infusion sets. Also, a more efficient drip chamber
adds an important advantage. Finally, increasing the tubing diameter
adds only minimal benefit.