D. Bienzle et al., COMPARISON OF 2 AUTOMATED MULTICHANNEL HEMATOLOGY ANALYZERS IN DOMESTIC-ANIMALS, Comparative haematology international, 4(3), 1994, pp. 162-166
Blood samples from 33 dogs, 28 cats, 24 horses and 25 cattle were anal
ysed in duplicate on the Coulter Counter S-Plus IV and the Baker 9000
multichannel haematology analysers. The precisions of each instrument
and the correlations between the instruments were evaluated. The preci
sions of both systems were good but, the Baker 9000 system showed more
variability between duplicate determinations for all parameters. The
Baker 9000 had better precision for white blood cell count, haemoglobi
n concentration, and MCV. Correlations between the two instruments wer
e excellent except for MCHC. Neither analyser reliably provided feline
platelet counts and the Baker 9000 also failed to report error messag
es for 8 of 10 apparently false platelet counts. The Coulter instrumen
t consistently reported higher haemoglobin concentrations and higher f
eline and lower canine white blood cell counts than the Baker 9000.