COMPARISON OF 2 AUTOMATED MULTICHANNEL HEMATOLOGY ANALYZERS IN DOMESTIC-ANIMALS

Citation
D. Bienzle et al., COMPARISON OF 2 AUTOMATED MULTICHANNEL HEMATOLOGY ANALYZERS IN DOMESTIC-ANIMALS, Comparative haematology international, 4(3), 1994, pp. 162-166
Citations number
5
Categorie Soggetti
Hematology
ISSN journal
09387714
Volume
4
Issue
3
Year of publication
1994
Pages
162 - 166
Database
ISI
SICI code
0938-7714(1994)4:3<162:CO2AMH>2.0.ZU;2-#
Abstract
Blood samples from 33 dogs, 28 cats, 24 horses and 25 cattle were anal ysed in duplicate on the Coulter Counter S-Plus IV and the Baker 9000 multichannel haematology analysers. The precisions of each instrument and the correlations between the instruments were evaluated. The preci sions of both systems were good but, the Baker 9000 system showed more variability between duplicate determinations for all parameters. The Baker 9000 had better precision for white blood cell count, haemoglobi n concentration, and MCV. Correlations between the two instruments wer e excellent except for MCHC. Neither analyser reliably provided feline platelet counts and the Baker 9000 also failed to report error messag es for 8 of 10 apparently false platelet counts. The Coulter instrumen t consistently reported higher haemoglobin concentrations and higher f eline and lower canine white blood cell counts than the Baker 9000.