Ce. Floyd et al., SCATTER-REDUCTION CHARACTERISTICS OF AN INFINITY-FOCUSED GRIDDED RADIOGRAPHIC CASSETTE, Investigative radiology, 29(9), 1994, pp. 852-855
RATIONALE AND OBJECTIVES. The scatter-reduction properties of a new in
finity-focused grid incorporated into a radiographic imaging cassette
were analyzed. METHODS. A polystyrene chest phantom was imaged using t
he cassette and bedside radiographic procedure. Scatter fractions were
measured using a beam-stop technique at several anatomically equivale
nt locations. The performance of this cassette also was evaluated as a
function of the orientation angle. RESULTS. The gridded cassette prov
ided a decrease in scatter fraction from 61% to 49% in the lung and fr
om 87% to 76% in the mediastinum. To obtain equivalent film density wh
en the grid was used, the exposure was increased by a factor of 3.0. W
hile there was a decrease in scatter clean-up as the film cassette was
tilted from the perpendicular, the grid performed well out to an angl
e of 10 degrees. CONCLUSION. The gridded cassette for portable radiogr
aphy provides scatter reduction with little sensitivity to alignment.
The availability of this device could improve scatter rejection, and t
herefore, contrast in portable radiographic imaging.