O. Benary et al., LIQUID IONIZATION CALORIMETRY WITH TIME-SAMPLED SIGNALS, Nuclear instruments & methods in physics research. Section A, Accelerators, spectrometers, detectors and associated equipment, 349(2-3), 1994, pp. 367-383
We present the results of a study of amplitude and timing measurements
made in a liquid krypton electromagnetic calorimeter, using multiply
sampled signals of the shaped waveform. The measurements were designed
to emulate the type of data that will be available from a calorimeter
operating at future hadron-hadron colliders with short (approximately
20 ns) spacing between bunch crossings. Data have been collected with
18 ns sample spacing on waveforms from individual calorimeter section
s with a shaping time of 40 ns and from 5 x 5 tower analog sums with a
shaping time of 50 ns. The amplitude was measured using the analog su
m signal, and the timing was measured using the signal from the indivi
dual sections. The data were processed using the method of optimal fil
tering, and a reduction in the noise of about a factor of two over tha
t for a single sample is seen when using multiple samples for determin
ing the amplitude. We find an energy resolution of 6.7%/square-root E,
in agreement with the resolution measured for the same calorimeter us
ing a single sample measured at the peak of the waveform. The timing r
esolution for a section of a calorimeter tower with deposited energy e
psilon can be expressed as square-root (c/epsilon)2 + sigma(cal)2, wit
h a value of c of 0.38 GeV ns for the front section (the first 6 radia
tion lengths) and 0.70 GeV ns for the back section, and a value of 0.1
5 ns for sigma(cal).