Kj. Weeks et al., CLINICAL USE OF A CONCOMITANT BOOST TECHNIQUE USING A GYPSUM COMPENSATOR, International journal of radiation oncology, biology, physics, 30(3), 1994, pp. 693-698
Citations number
12
Categorie Soggetti
Oncology,"Radiology,Nuclear Medicine & Medical Imaging
Purpose: To develop a clinical procedure to treat field within a field
(concomitant boost) portals with a single compensated field. Methods
and Materials: An ordinary manual cerrobend block former was used to p
roduce styrofoam molds from simulator film data. A special gypsum comp
ound was poured into the molds. The compensator block is independently
mounted to the treatment machine via a custom-made compensator holder
. Results: Measurements confirm that the inhomogeneous dose distributi
on has been reliably delivered via this technique. The accuracy of pla
cement of the high dose region is sufficient for clinical use. Conclus
ion: The procedure enables the concomitant boost effect to be easily i
mplemented in the clinic without increasing clinical setup time.