DOPPLER ANALYSIS OF HEMODIALYSIS GRAFTS - RESISTIVE INDEX AS A PREDICTOR OF STENOSIS

Citation
De. Baird et al., DOPPLER ANALYSIS OF HEMODIALYSIS GRAFTS - RESISTIVE INDEX AS A PREDICTOR OF STENOSIS, Journal of ultrasound in medicine, 13(10), 1994, pp. 791-796
Citations number
17
Categorie Soggetti
Acoustics,"Radiology,Nuclear Medicine & Medical Imaging
ISSN journal
02784297
Volume
13
Issue
10
Year of publication
1994
Pages
791 - 796
Database
ISI
SICI code
0278-4297(1994)13:10<791:DAOHG->2.0.ZU;2-V
Abstract
A series of six hemodialysis vascular access patients was evaluated wi th duplex Doppler sonography and angiography to identify Doppler sonog raphic parameters that are predictors of venous outflow stenosis. This preliminary series indicates that no venous outflow stenosis was pres ent angiographically when the resistive index gradient across the veno us anastomosis is less than or equal to zero and a significant venous outflow stenosis (50 to 100%) was present when the resistive index gra dient is 0.15 or greater (P = 0.002, r = 0.963). Doppler acquisition o f resistive index gradients across regions of venous outflow is a prom ising, noninvasive means of rapid radiologic evaluation of hemodialysi s graft patients, especially when the need for angiography is equivoca l.