C. Lefez et al., INTRA-BRONCHIAL MIGRATION AND EXTRACTION OF A FRAGMENT OF AN ACCUFIX J-SHAPED ATRIAL-PACING CATHETER, Archives des maladies du coeur et des vaisseaux, 90(2), 1997, pp. 291-293
The authors report a case of exteriorisation and migration of a fragme
nt of a J-shaped atrial Accufix Telectronix (Stimarec Class IV) respon
sable for a bronchial penetration without clinical repercussion. A sys
tematic control chest X-ray detected this complication. The position o
f the metallic fragment was determined by chest CT scan and bronchial
fibroscopy between the inferior right lobar bronches and artery confir
ming bronchial perforation. Rigid bronchoscopy with direct visualisati
on enabled extraction of the foreign body without complications.