MEASUREMENTS OF ISOBARIC HEAT-CAPACITIES OF GASES FROM 323.15-K TO 573.15-K UP TO 30 MPA

Citation
L. Dordain et al., MEASUREMENTS OF ISOBARIC HEAT-CAPACITIES OF GASES FROM 323.15-K TO 573.15-K UP TO 30 MPA, Review of scientific instruments, 65(10), 1994, pp. 3263-3267
Citations number
15
Categorie Soggetti
Physics, Applied","Instument & Instrumentation
ISSN journal
00346748
Volume
65
Issue
10
Year of publication
1994
Pages
3263 - 3267
Database
ISI
SICI code
0034-6748(1994)65:10<3263:MOIHOG>2.0.ZU;2-X
Abstract
A Setaram C-80 calorimeter has been modified in order to measure heat capacities at constant pressure of gases from ambient temperature to 5 73 K and at pressures up to 30 MPa. Special cells have been designed a nd an additional device has been incorporated into the calorimeter to better control the temperature. The calorimetric technique and the cal ibration method are described. Experimental heat capacities of CO2 in the temperature range 323.15-423.15 K at pressures up to 25 MPa are re ported and compared with the best literature data.