CAN A DRAWOVER VAPORIZER BE A PUSHOVER

Citation
Jc. Taylor et J. Restall, CAN A DRAWOVER VAPORIZER BE A PUSHOVER, Anaesthesia, 49(10), 1994, pp. 892-894
Citations number
5
Categorie Soggetti
Anesthesiology
Journal title
ISSN journal
00032409
Volume
49
Issue
10
Year of publication
1994
Pages
892 - 894
Database
ISI
SICI code
0003-2409(1994)49:10<892:CADVBA>2.0.ZU;2-E
Abstract
Bench testing was carried out to establish whether the vapour output f rom an OMV50 vaporizer, as used in the Triservice apparatus, differs a ccording to whether the carrier gas is either drawn or pushed through the vaporizer. Results show that the differences in output concentrati on between the two modes were clinically insignificant.