CARBON-PASTE BIOSENSORS ARRAY FOR LONG-TERM GLUCOSE MEASUREMENT

Authors
Citation
J. Kulys et He. Hansen, CARBON-PASTE BIOSENSORS ARRAY FOR LONG-TERM GLUCOSE MEASUREMENT, Biosensors & bioelectronics, 9(7), 1994, pp. 491-500
Citations number
30
Categorie Soggetti
Biothechnology & Applied Migrobiology",Biophysics
Journal title
ISSN journal
09565663
Volume
9
Issue
7
Year of publication
1994
Pages
491 - 500
Database
ISI
SICI code
0956-5663(1994)9:7<491:CBAFLG>2.0.ZU;2-Z
Abstract
An amperometric glucose biosensor based on a new mediator and carbon p aste acts at 0.08 V vs. an Ag/AgCl electrode in situ and exhibits a li near response up to 16 mM of glucose. The measurements are performed o n freshly renovated electrode surfaces, and every sensor can be used f or 100 measurements. The long-term stable biosensors were obtained usi ng stabilized glucose oxidase and an array of biosensors. The applicat ion of a chemometric method permits exploitation of the biosensor duri ng a period of more than 166 days at 35-degrees-C and relative humidit y 75%. The relative error of chemometric prediction was less than 5.4% , and the response changes less than 0.15% per week.