Cg. Xie et al., EXPERIMENTAL EVALUATION OF CAPACITANCE TOMOGRAPHIC FLOW IMAGING-SYSTEMS USING PHYSICAL MODELS, IEE proceedings. Circuits, devices and systems, 141(5), 1994, pp. 357-368
An experimental method for evaluating the performance of capacitance t
omographic flow imaging systems is described. Criteria are defined to
describe the performance of these systems, namely spatial and permitti
vity resolution, accuracy (system errors) and signal to noise ratio. S
tatic physical models simulating typical flow distribution patterns ar
e used as standard test objects and the criteria are quantified by com
paring the reconstructed images with the standards. Systems with 8 and
12 electrodes have been tested and the results compared to show the i
mprovement obtained with the increased number of electrodes.