ADSORPTION OF HUMAN RECOMBINANT ERYTHROPOIETIN ON DIALYSIS MEMBRANES IN-VITRO
Citation
H. Mori et al., ADSORPTION OF HUMAN RECOMBINANT ERYTHROPOIETIN ON DIALYSIS MEMBRANES IN-VITRO, Artificial organs, 18(10), 1994, pp. 725-728
Categorie Soggetti
Engineering, Biomedical
SICI code
0160-564X(1994)18:10<725:AOHREO>2.0.ZU;2-9
Abstract
The adsorptive characteristics of 5 dialysis membranes for recombinant
human erythropoietin (EPO) were studied in vitro in a closed circuit
system. For 120 min, EPO added with bovine serum was significantly ads
orbed by polymethylmetacrylate (PMMA) and polyacrylonitrile (PAN) memb
ranes but not by Cuprophan, ethylene vinyl alcohol (EVAL), or polysulf
one (PS) membranes. In addition the EPO adsorptive rate, as well as th
e of beta2-microglobulin (beta2-MG), was greater with a PMMA membrane
than with a PAN membrane. EPO was not detected in the ultrafiltrate at
15 min with 5 membranes. These results indicate that EPO was eliminat
ed by membrane adsorption only with some dialysis membranes.