DEDUCTIVE ELECTROCHEMICAL DETECTION FOR CAPILLARY ELECTROPHORESIS

Citation
Ma. Malone et al., DEDUCTIVE ELECTROCHEMICAL DETECTION FOR CAPILLARY ELECTROPHORESIS, Analytical chemistry, 66(21), 1994, pp. 3782-3787
Citations number
30
Categorie Soggetti
Chemistry Analytical
Journal title
ISSN journal
00032700
Volume
66
Issue
21
Year of publication
1994
Pages
3782 - 3787
Database
ISI
SICI code
0003-2700(1994)66:21<3782:DEDFCE>2.0.ZU;2-D
Abstract
A system has been developed for reductive electrochemical detection wi th capillary electrophoresis. The instrumental setup is much simpler t han that of reductive liquid chromatography/ electrochemistry (LCEC). The new method also possesses major advantages over LC-based methods i n terms of a less complicated deoxygenation system and shorter deoxyge nation times. Concentration detection Limits were similar to those rep orted for reductive LCEC. Mass detection limits for anthraquinone-2-ca rboxylic acid and (dinitrophenyl)-gamma-aminobutyric acid were 1.3 and 1.6 fmol, respectively, at a S/N of 2. The reproducibility of the sig nal for a 9.8-fmol anthraquinone-2-carboxylic acid injection yielded a relative standard deviation of 6.2%. The high selectivity of reductiv e electrochemical detection was demonstrated by the detection of mitom ycin C directly in human serum without prior extraction procedures.