X-RAY POSITION RESOLUTION IN PROPORTIONAL CHAMBERS IN THE REGION OF 100 MU (FWHM) ABOVE THE XENON K-EDGE

Citation
Gc. Smith et al., X-RAY POSITION RESOLUTION IN PROPORTIONAL CHAMBERS IN THE REGION OF 100 MU (FWHM) ABOVE THE XENON K-EDGE, Nuclear instruments & methods in physics research. Section A, Accelerators, spectrometers, detectors and associated equipment, 350(3), 1994, pp. 621-623
Citations number
7
Categorie Soggetti
Nuclear Sciences & Tecnology","Physics, Particles & Fields","Instument & Instrumentation",Spectroscopy
ISSN journal
01689002
Volume
350
Issue
3
Year of publication
1994
Pages
621 - 623
Database
ISI
SICI code
0168-9002(1994)350:3<621:XPRIPC>2.0.ZU;2-V
Abstract
A brief discussion is presented of the X-ray absorption process in xen on above its K-edge of 34.6 keV. Experimental measurements from a one- dimensional proportional chamber, operating with a gas mixture of Xe/1 0%CO2, show that position resolution in the region of 100 mum FWHM can be achieved for X-rays with energy just above the K-edge.