A rapid, simple and accurate HPLC method is presented for the determin
ation of nadoxolol in human plasma. Nadoxolol from plasma was successf
ully purified using an Adsorbex column. The samples were chromatograph
ed on a LiChrosorb RP-18 (10 mu m) column with methanol-acetonitrile-p
hosphate buffer (pH 3.3) (70:20:10) as the mobile phase. Detection was
carried out at 254 nm. The method was tested for linearity (from 5 to
25 mu g/ml), recovery (85%) and precision (C.V. = 4.5%).