PATIENT RADIATION-DOSES FROM ENTEROCLYSIS EXAMINATIONS

Citation
D. Hart et al., PATIENT RADIATION-DOSES FROM ENTEROCLYSIS EXAMINATIONS, British journal of radiology, 67(802), 1994, pp. 997-1000
Citations number
13
Categorie Soggetti
Radiology,Nuclear Medicine & Medical Imaging
Journal title
British journal of radiology
ISSN journal
00071285 → ACNP
Volume
67
Issue
802
Year of publication
1994
Pages
997 - 1000
Database
ISI
SICI code
Abstract
Data relating to patient dose have been acquired for enteroclysis exam inations (small bowel enemas) performed at the John Radcliffe Hospital , Oxford, on 23 adult patients. Dose-area products, fluoroscopy times and the number of radiographs taken are used to compare the examinatio n procedure at the John Radcliffe Hospital with enteroclysis and bariu m follow-throughs performed elsewhere. The mean dose-area product for the 23 examinations was 6.8 Gy cm(2) and the mean effective dose was e stimated to be 1.5 mSv. These doses are intermediate between those ari sing from barium meals and barium enemas performed in the same room.