EFFECT OF PULSATING FLOW ON CORIOLIS MASS FLOWMETERS

Authors
Citation
G. Vetter et S. Notzon, EFFECT OF PULSATING FLOW ON CORIOLIS MASS FLOWMETERS, FLOW MEASUREMENT AND INSTRUMENTATION, 5(4), 1994, pp. 263-273
Citations number
NO
Categorie Soggetti
Instument & Instrumentation","Engineering, Mechanical
ISSN journal
09555986
Volume
5
Issue
4
Year of publication
1994
Pages
263 - 273
Database
ISI
SICI code
0955-5986(1994)5:4<263:EOPFOC>2.0.ZU;2-M
Abstract
The last ten years have seen an increasing use of Coriolis flowmeters in the measurement of small flows of liquids. The performance of Corio lis meters under pulsating flow conditions was not well known, however . This paper presents some investigations on such flowmeters in the pr esence of monofrequent and polyfrequent pulsating flows. The linearity of the flow characteristics is checked mathematically and the mechani cal resonances of the measuring tubes are determined. Two U-tube type Coriolis meters are tested and the measuring error resulting from the pulsation of the flow is examined. The investigations are carried out with monofrequent pulsation by a special test rig flow and with geared and piston type pumps (polyfrequent excitation).