A METHOD TO DETERMINE A TRANSMISSION CORRECTION FOR ELECTRON SPECTROMETERS USING SYNCHROTRON-RADIATION

Citation
J. Jauhiainen et al., A METHOD TO DETERMINE A TRANSMISSION CORRECTION FOR ELECTRON SPECTROMETERS USING SYNCHROTRON-RADIATION, Journal of electron spectroscopy and related phenomena, 69(3), 1994, pp. 181-187
Citations number
12
Categorie Soggetti
Spectroscopy
ISSN journal
03682048
Volume
69
Issue
3
Year of publication
1994
Pages
181 - 187
Database
ISI
SICI code
0368-2048(1994)69:3<181:AMTDAT>2.0.ZU;2-O
Abstract
The transmission function of a hemispherical electron spectrometer, th e SES-144 (Osborne et al., Synchrotron Radiat. News, 7 (1) (1994) 25), has been determined for pass energies of 10, 20 and 50 eV in the kine tic energy range 7-60 eV. The method is based on the measurement of th e intensity ratio between photoelectron lines with varying kinetic ene rgies and the corresponding constant kinetic energy Auger lines using different photon energies of synchrotron radiation. With the aid of th e obtained transmission curves, the ''standard'' Auger line intensitie s of the Xe N4,5OO Auger spectrum have been determined. Using these in tensities as reference values, the transmission curves of other electr on spectrometers can conveniently be found.