R. Livi et al., SIMULTANEOUS AND SEQUENTIAL SAME-ARM MEASUREMENTS IN THE VALIDATION STUDIES OF AUTOMATED BLOOD-PRESSURE MEASURING DEVICES, American journal of hypertension, 9(12), 1996, pp. 1228-1231
The oscillometric ambulatory blood pressure recorder Daypress 500 was
validated according to the British Hypertension Society protocol. Both
sequential and simultaneous measurements were used. Multiple regressi
on analysis demonstrated a significant influence of subject pulse pres
sure and arm circumference on device-observer systolic pressure differ
ences. Differences between observer consecutive readings were inversel
y related to heart rate. Device and observer blood pressure readings w
ere closer at simultaneous than at sequential measurements. However, b
oth kinds of measurement led to the same final evaluation (A for diast
olic and B for systolic blood pressure), provided that the appropriate
grading criteria were applied for each method. (C) 1996 American Jour
nal of Hypertension, Ltd.