SIMULTANEOUS AND SEQUENTIAL SAME-ARM MEASUREMENTS IN THE VALIDATION STUDIES OF AUTOMATED BLOOD-PRESSURE MEASURING DEVICES

Citation
R. Livi et al., SIMULTANEOUS AND SEQUENTIAL SAME-ARM MEASUREMENTS IN THE VALIDATION STUDIES OF AUTOMATED BLOOD-PRESSURE MEASURING DEVICES, American journal of hypertension, 9(12), 1996, pp. 1228-1231
Citations number
9
Categorie Soggetti
Cardiac & Cardiovascular System
ISSN journal
08957061
Volume
9
Issue
12
Year of publication
1996
Part
1
Pages
1228 - 1231
Database
ISI
SICI code
0895-7061(1996)9:12<1228:SASSMI>2.0.ZU;2-G
Abstract
The oscillometric ambulatory blood pressure recorder Daypress 500 was validated according to the British Hypertension Society protocol. Both sequential and simultaneous measurements were used. Multiple regressi on analysis demonstrated a significant influence of subject pulse pres sure and arm circumference on device-observer systolic pressure differ ences. Differences between observer consecutive readings were inversel y related to heart rate. Device and observer blood pressure readings w ere closer at simultaneous than at sequential measurements. However, b oth kinds of measurement led to the same final evaluation (A for diast olic and B for systolic blood pressure), provided that the appropriate grading criteria were applied for each method. (C) 1996 American Jour nal of Hypertension, Ltd.