A TESTING OF THE FLOW-INJECTION TECHNIQUE ON THE SEQUENTIAL ICP-SPECTROMETER 3510-ICP

Authors
Citation
S. Gergely et M. Cakrt, A TESTING OF THE FLOW-INJECTION TECHNIQUE ON THE SEQUENTIAL ICP-SPECTROMETER 3510-ICP, Journal of analytical chemistry, 49(10), 1994, pp. 962-965
Citations number
6
Categorie Soggetti
Chemistry Analytical
ISSN journal
10619348
Volume
49
Issue
10
Year of publication
1994
Pages
962 - 965
Database
ISI
SICI code
1061-9348(1994)49:10<962:ATOTFT>2.0.ZU;2-P
Abstract
The flow injection technique (FI) was tested in order to compare the a nalytical performance of this technique with the conventional one. The statistical parameters of the calibration curves for conventional and flow injection technique were found to be comparable. Determination l imits obtained by the Fl technique were about 1 magnitude of order wor se then those obtained by the conventional one. The signal evaluation of the transient signal is discussed. The influence of NaCl and nitric acid upon the precision and accuracy of the analysis for Ca, Co and Z n was also investigated.