A NEW COAXIAL NEEDLE SYSTEM, HEPATIC-ARTERY TARGETING WIRE, AND BIPLANE FLUOROSCOPY TO INCREASE SAFETY AND EFFICACY OF TIPS

Citation
O. Matsui et al., A NEW COAXIAL NEEDLE SYSTEM, HEPATIC-ARTERY TARGETING WIRE, AND BIPLANE FLUOROSCOPY TO INCREASE SAFETY AND EFFICACY OF TIPS, Cardiovascular and interventional radiology, 17(6), 1994, pp. 343-346
Citations number
NO
Categorie Soggetti
Radiology,Nuclear Medicine & Medical Imaging
ISSN journal
01741551
Volume
17
Issue
6
Year of publication
1994
Pages
343 - 346
Database
ISI
SICI code
0174-1551(1994)17:6<343:ANCNSH>2.0.ZU;2-U
Abstract
To improve safety and efficacy of the transjugular intrahepatic portos ystemic shunt (TIPS) procedure, we introduced a new, thin-needle (21-g auge long PTC needle) puncture technique using biplane fluoroscopy and targeting of a guidewire tip in the right hepatic artery. After punct ure of the right portal vein, a 0.016-inch guidewire was inserted into the portal vein, followed by a 4 Fr dilator. The 4 Fr dilator allowed introduction of a 0.035-inch working guidewire. We successfully perfo rmed TIPS in seven patients with postnecrotic cirrhosis using this tec hnique and encountered no technical difficulties or complications.