LOW-TEMPERATURE 2-AXIS GONIOMETER WITH ACCURATE TEMPERATURE CONTROL

Citation
R. Herzog et Je. Evetts, LOW-TEMPERATURE 2-AXIS GONIOMETER WITH ACCURATE TEMPERATURE CONTROL, Review of scientific instruments, 65(11), 1994, pp. 3574-3576
Citations number
4
Categorie Soggetti
Physics, Applied","Instument & Instrumentation
ISSN journal
00346748
Volume
65
Issue
11
Year of publication
1994
Pages
3574 - 3576
Database
ISI
SICI code
0034-6748(1994)65:11<3574:L2GWAT>2.0.ZU;2-M
Abstract
We have designed a probe to rotate a sample around two independent axe s in a magnetic field of up to 8 T and at temperatures varying from 3 K to room temperature. The sample is in very good thermal contact with a thermometer and a regulated heater, enabling precise temperature co ntrol. Pogo pins enable quick electrical connections to the sample. Tw o computer-controlled stepper motors rotate the sample around the two independent axes. This probe was primarily designed to investigate the anisotropy of the critical currents of superconducting thin films in a magnetic field.