DETERMINATION OF COPPER BY ZEOLITE MOLECULAR-SIEVE MODIFIED ELECTRODE

Citation
B. Chen et al., DETERMINATION OF COPPER BY ZEOLITE MOLECULAR-SIEVE MODIFIED ELECTRODE, Electrochimica acta, 42(4), 1997, pp. 595-604
Citations number
22
Categorie Soggetti
Electrochemistry
Journal title
ISSN journal
00134686
Volume
42
Issue
4
Year of publication
1997
Pages
595 - 604
Database
ISI
SICI code
0013-4686(1997)42:4<595:DOCBZM>2.0.ZU;2-X
Abstract
Chemically modified carbon paste electrodes incorporated with zeolite molecular sieves were constructed as sensor electrodes for voltammetri c determination of trace copper in aqueous solution. Square wave volta mmetric techniques were used to quantify Cu(II) in aqueous solution. T he copper response was characterized with respect to electrode composi tion, cleaning solution, Cu(II) concentration, preconcentration time, pH, and other variables. The peak current was found to be linearly rel ated to the concentration of Cu(II) in the range of 0.3-1.0 mg l(-1) w ith a correlation coefficient of 0.9992. The electrodes were electroch emically sensitive to copper in ammonia buffer solution with a detecti on limit of 0.074 mg l(-1). Copyright (C) 1996 Elsevier Science Ltd