Principles and methods for scanning electrode quartz crystal analysis
(SEQCA) as a technique for mapping distribution of mass and measuring
interfacial rheology are described, In SEQCA, quartz crystal resonates
in a region between a normal electrode and a microelectrode scanned o
ver the opposite side of the normal electrode. The surface of the quar
tz crystal plate just below the scanning electrode is monitored (overs
canning mode), A prototype system for the scanning electrode quartz cr
ystal analysis is constructed. The dependence of the resonance propert
ies on the location of the scanning electrode is studied. As a result,
it was found that the qualitative mapping of the mass distribution is
possible in the lateral resolution of 1 (9 MHz quartz) to 2 (5 MHz) m
m, Viscoelastic properties of the load on the quartz surface are also
measured qualitatively.