ALL-SOLID-STATE HYDROGEN ION-SELECTIVE ELECTRODE BASED ON A CONDUCTING POLY(PYRROLE) SOLID CONTACT

Citation
A. Michalska et al., ALL-SOLID-STATE HYDROGEN ION-SELECTIVE ELECTRODE BASED ON A CONDUCTING POLY(PYRROLE) SOLID CONTACT, Analyst, 119(11), 1994, pp. 2417-2420
Citations number
24
Categorie Soggetti
Chemistry Analytical
Journal title
ISSN journal
00032654
Volume
119
Issue
11
Year of publication
1994
Pages
2417 - 2420
Database
ISI
SICI code
0003-2654(1994)119:11<2417:AHIEBO>2.0.ZU;2-0
Abstract
A potentiometric ion-selective electrode based on a solid conducting p oly(pyrrole) contact and neutral carrier tri-n-dodecylamine membrane w as developed. The effect of the poly(pyrrole) doping anion on the char acteristic of the sensor is presented and discussed. The ion-selective H+ sensitive electrode with a solid poly(pyrrole) contact is characte rized by having a good stability over both short and long periods of t ime. Selectivity coefficients for this electrode are slightly better t han those for internal solution electrodes. Stability is also improved especially when compared with coated rr;ire type electrodes. Response s are fast and nearly Nernstian within the pH range 2-12. All solid-st ate sensors with plastic tri-n-dodecylamine were used for pH determina tions in blood serum.