CHOICE OF MOVING TARGET FOR A STRING PHANTOM .2. ON THE PERFORMANCE TESTING OF DOPPLER ULTRASOUND SYSTEMS

Authors
Citation
Pr. Hoskins, CHOICE OF MOVING TARGET FOR A STRING PHANTOM .2. ON THE PERFORMANCE TESTING OF DOPPLER ULTRASOUND SYSTEMS, Ultrasound in medicine & biology, 20(8), 1994, pp. 781-789
Citations number
8
Categorie Soggetti
Radiology,Nuclear Medicine & Medical Imaging",Acoustics
ISSN journal
03015629
Volume
20
Issue
8
Year of publication
1994
Pages
781 - 789
Database
ISI
SICI code
0301-5629(1994)20:8<781:COMTFA>2.0.ZU;2-H
Abstract
A string phantom was used to test a duplex system using several filame nts: braided silk, nylon and O-ring rubber. The estimated pulsatility index was identical for ah filaments tested. Range gate registration m easurements were identical for all filaments tested. Maximum velocity was underestimated in the region of a strong angular backscatter peak this was the case for several of the braided silk filaments. This effe ct was not observed for O-ring rubber. The paper concluded that O-ring rubber is a suitable filament for the testing of several key Doppler quantities. Recalibration of the string phantom velocity indicator is necessary due to the large diameter of the O-ring rubber.