ION-SELECTIVE MEMBRANE-ELECTRODE FOR IMIP RAMINE - CONSTRUCTION, ELECTROANALYTICAL CHARACTERIZATION AND PHARMACEUTICAL APPLICATION

Citation
Ri. Stefan et al., ION-SELECTIVE MEMBRANE-ELECTRODE FOR IMIP RAMINE - CONSTRUCTION, ELECTROANALYTICAL CHARACTERIZATION AND PHARMACEUTICAL APPLICATION, Revista de chimie, 45(10), 1994, pp. 903-907
Citations number
14
Categorie Soggetti
Chemistry
Journal title
ISSN journal
00347752
Volume
45
Issue
10
Year of publication
1994
Pages
903 - 907
Database
ISI
SICI code
0034-7752(1994)45:10<903:IMFIR->2.0.ZU;2-I
Abstract
The paper presents the construction and the performance characteristic s of a new membrane electrode selective to imipramine cation drug. The construction is based on the used of the ion-pair association of imip ramine hydrochloride and dinonylnaphthalenesulphonic acid embedded in a PVC matrix. The electrode exhibits a near-Nernstian response in the range 10(-2) - 10(-5) mol 1(-1) and a low detection limit of about 5x1 0(-6) mol 1(-1). It was studied its selectivity over a series of the p otentially interfering species. The electrode response was independent of pH in the range 2.5-5.0. The electrode can be used successfully fo r the determination of imipramine hydrochloride in pharmaceuticals wit hout any prior separation, with an average error of less than 1.00%.