A. Pilo et al., EVALUATION OF PRECISION AND RELATIVE ACCURACY OF A NEW AUTOMATED LUMINESCENCE IMMUNOASSAY FOR TSH AND FT4, Journal of clinical immunoassay, 17(3), 1994, pp. 176-182
The chemiluminescence immunoasssay system LIA-mat S 300 for TSH and fT
4, developed by Byk-Sangtec Diagnostica (Dietzenbach, Germany), has be
en evaluated in a multicenter collaborative study among 5 laboratories
, using control materials previously circulated in a national External
Quality Assessment (EQA) program. LIA-mat S 300 is an open, fully aut
omated system in which all the analytical steps (pipetting of samples
and reagents, mixing, incubation, washing, luminescent measurement and
data reduction) are controlled by a programmable computer. TSH is mea
sured by a noncompetitive technique using two different monoclonal ant
ibodies. In fT4 assay the free hormone in the sample and a fixed amoun
t of T4 conjugate -coated on the tube-compete for binding to a monoclo
nal anti T4 antibody labeled with an isoluminol derivative. The workin
g range of LIA-mat TSH (computed from the precision profile, as the co
ncentration range measurable with a within-assay precision better than
10 CV%) extended from 0.09 to 80 mu IU/mL; the sensitivity was 0.017
mu IU/ mL. The mean inprecision in the concentration range 0.25-43 mu
IU/mL, were 6.0 CV% (between-assay) and 8.7 CV% (between-lab). The com
parison of the results from LIA-mat and those obtained by other method
s/kits on EQA control samples indicated a satisfactory agreement. The
working range of fT4 extended from 0.37 to 8.96 ng/dL and the sensitiv
ity was 0.07 ng/dL. The mean between-assay variability was 8.8 CV% for
samples at concentration below 1.0 ng/dL and 5.7 CV% for higher sampl
es; the between-lab variability was 10.2 CV%. The relative accuracy of
LIA-mat fT4 has been evaluated by comparison with the mean of the met
hods/kits more used in the EQA; we observed that the agreement between
the different methods, including LIA-mat fT4, is satisfactory for con
trol samples in the normal and in the hyperthyroid range. Moreover the
measurement of LIA-mat fT4, as theoretically predicted, is not affect
ed by dilution, it produces a result in the lower normal range for a p
regnancy control pool and finds in a FDH (Familial Disalbuminemic Hype
rthyroxinemia) pool a value in the borderline normal-hyperthyroid rang
e.