G. Lespes et al., ICP-MS, A REVIEW - APPLICATIONS TO SPECIATION OF TRACE-ELEMENTS IN AQUATIC AND BIOLOGICAL ENVIRONMENTS, Environmental technology, 13(3), 1992, pp. 207-220
The speciation of organometallic compounds using Inductively Coupled P
lasma Mass Spectrometry (ICP-MS) as a detector is reviewed. Two differ
ent methodologies have been developped: separative extraction and liqu
id or gas chromatography, coupled with ICP-MS. The main experimental c
onditions described in each paper are listed. Selectivity of the metho
ds, detection limits, chromatographic parameters, experimental conditi
ons of ICP-MS and interferences are discussed.