Aj. Vanhoute et Pcm. Bartels, COMPARATIVE-EVALUATION OF THE PHARMACIA CAP SYSTEM AND THE DPC ALASTAT SYSTEM FOR INVITRO DETECTION OF ALLERGEN-SPECIFIC IGE WITH THE SKIN PRICK TEST, European journal of clinical chemistry and clinical biochemistry, 30(2), 1992, pp. 101-105
Allergen-specific IgE concentrations for 6 inhalant allergens were com
pared in 54 subjects, using the Pharmacia CAP System and the DPC AlaST
AT System. Similar results were obtained in 91.6% of the tests. In 8%
of the tests a positive score was obtained for the Pharmacia CAP Syste
m whereas a negative result was found by use of the DPC AlaSTAT System
. A negative Pharmacia CAP System result combined with a positive DPC
AlaSTAT System result occurred only in 0.4% of the tests. Sensitivity
and specificity was established for the Pharmacia CAP System and the D
PC AlaSTAT System by comparison with results of the skin prick test. R
esults obtained with the Pharmacia CAP System were found to be of slig
htly higher sensitivity than those obtained by DPC AlaSTAT System and
of similar or slightly lower specificity.