Xl. Lin et K. Heydorn, EFFECT OF CASCADE COINCIDENCES ON THE EFFICIENCY CALIBRATION OF A GAMMA-X DETECTOR, Journal of radioanalytical and nuclear chemistry, 169(2), 1993, pp. 419-427
The sensitivity on n-type gamma-X detectors for low-energy X- and gamm
a-rays calls for coincidence corrections in the efficiency calibration
that do not apply to the calibration of p-type detectors. Corrections
were calculated for the effect of cascade coincidences between gamma-
rays, X-rays, annihilation radiation, and bremsstrahlung, for 15 radio
nuclides frequently used for efficiency calibration. Experimental resu
lts are presented for a gamma-X detector with 37% relative efficiency
at distances from 0.9 to 17.5 cm. After coincidence correction smooth
efficiency curves were found for the energy range 12 to 2750 keV, even
for the position closest to the detector.