SCINTILLATION ELECTROMAGNETIC CALORIMETER WITH OPTICAL-FIBER LIGHT OUTPUT

Citation
Sv. Afanasyev et al., SCINTILLATION ELECTROMAGNETIC CALORIMETER WITH OPTICAL-FIBER LIGHT OUTPUT, Instruments and experimental techniques, 35(5), 1992, pp. 793-799
Citations number
9
ISSN journal
00204412
Volume
35
Issue
5
Year of publication
1992
Part
1
Pages
793 - 799
Database
ISI
SICI code
0020-4412(1992)35:5<793:SECWOL>2.0.ZU;2-B
Abstract
This paper describes a scintillation calorimeter with an optical-fiber light output of 47 lead and 47 scintillation plates of sizes 2 x 70 x 70 and 5 x 70 x 70 mm, correspondingly. In total the unit is 17.3 rad iation-lengths thick. The unit was tested using electron beams with en ergies from 0.5 to 4.5 GeV. The energy resolution of the calorimeter i s sigma/E = (0.114 +/- 0.005)/square-root E, where E is expressed in g iga-electronvolts, and the coordinate resolution is at least 4 mm. The muon-electron separation coefficient is 400.