To study the interior design of model 6702 and 6711 iodine-125 seeds c
ontact autoradiographs were performed using mammography film. Improved
resolution was obtained using a pin-hole camera with a hole of 0.1 mm
x 0.1 mm. With these techniques, qualitative determination of the rel
ative activity distribution within each seed was possible. The number
of the activated resin spheres and the positions of the centers of the
se spheres can be exactly determined. A model calculation shows, that
variations in the arrangement of the activated spheres within a seed h
ave a moderate influence on the dose distribution at source distances
below 10 mm. Knowing the exact source configuration may be useful when
comparing dose calculations with measured data for model 6702 I-125 s
eeds which are currently employed in ophthalmic plaque and implant the
rapy of other tumors.