TRANSFER VOLUME ACCURACY - CALIBRATED WIRE AND PLASTIC LOOPS

Citation
Gr. Miller et Jl. Perry, TRANSFER VOLUME ACCURACY - CALIBRATED WIRE AND PLASTIC LOOPS, Laboratory medicine, 24(5), 1993, pp. 278-280
Citations number
NO
Journal title
ISSN journal
00075027
Volume
24
Issue
5
Year of publication
1993
Pages
278 - 280
Database
ISI
SICI code
0007-5027(1993)24:5<278:TVA-CW>2.0.ZU;2-K
Abstract
The urine volume delivered via the calibrated loop technique for quant itating microorganisms can vary by as much as +/-50%. This discrepancy results from investigator technique differences, as well as variation s in wire gauges and the internal diameters of loops. Using the Evans blue dye procedure, we tested calibrated twisted-wire loops and dispos able plastic loops and compared them with standard platinum loops.