Cm. Ma et al., WALL-CORRECTION AND ABSORBED-DOSE CONVERSION FACTORS FOR FRICKE DOSIMETRY - MONTE-CARLO CALCULATIONS AND MEASUREMENTS, Medical physics, 20(2), 1993, pp. 283-292
For megavoltage radiotherapy photon beams, EGS4 Monte Carlo calculatio
ns show, and experimental measurements confirm with an accuracy of 0.2
%, that glass or quartz-walled vials used in Fricke dosimetry increase
the dose in the Fricke solution. This is mainly caused by increased e
lectron scattering from the glass which increases the dose to the Fric
ke solution. The dose perturbation is shown to vary from nothing in a
Co-60 beam up to 2% in a 24-MV beam. For plastic vials of similar shap
es, calculations demonstrate that the effect is in the opposite direct
ion and even at high energies it is much less (0.2% to 0.5%).