HOW ACCURATELY CAN EGS4 PRESTA CALCULATE ION-CHAMBER RESPONSE/

Authors
Citation
Dwo. Rogers, HOW ACCURATELY CAN EGS4 PRESTA CALCULATE ION-CHAMBER RESPONSE/, Medical physics, 20(2), 1993, pp. 319-324
Citations number
19
Categorie Soggetti
Radiology,Nuclear Medicine & Medical Imaging
Journal title
ISSN journal
00942405
Volume
20
Issue
2
Year of publication
1993
Part
1
Pages
319 - 324
Database
ISI
SICI code
0094-2405(1993)20:2<319:HACEPC>2.0.ZU;2-K
Abstract
Ion-chamber responses are calculated for graphite, PMMA, and aluminum- walled ion chambers free-in-air in Co-60 beams and 200-keV beams for a graphite chamber. The EGS4 Monte Carlo system is used with various el ectron step-size algorithms, in particular the PRESTA algorithm, the m uch simpler ESTEPE constraint on the energy loss per step, and in comb ination. Contrary to previous reports, it is found that there are vari ations in the calculated ion chamber response of up to 3% in Co-60 bea ms and up to 8% in the 200-keV beam. It is recommended that all ion ch amber calculations be done with the PRESTA algorithm plus an additiona l constraint on the energy losses per step of 1% (or less for higher-Z materials). The systematic uncertainty in the calculations for Co-60 beams is found to be 1% (1sigma) and somewhat higher in 200-keV beams because of sensitivity to transport parameters such as AE and ECUT.