RARE-EARTH SCATTER FRACTIONS IN CHEST RADIOGRAPHY

Citation
Tj. Petrone et Kd. Steidley, RARE-EARTH SCATTER FRACTIONS IN CHEST RADIOGRAPHY, Medical physics, 20(2), 1993, pp. 475-479
Citations number
12
Categorie Soggetti
Radiology,Nuclear Medicine & Medical Imaging
Journal title
ISSN journal
00942405
Volume
20
Issue
2
Year of publication
1993
Part
1
Pages
475 - 479
Database
ISI
SICI code
0094-2405(1993)20:2<475:RSFICR>2.0.ZU;2-5
Abstract
A beam stop technique was used to measure the densitometric scatter fr actions under three regions of a humanoid chest phantom utilizing LaOB r and Gd2O2S screens. For these receptors, the scatter fractions under the lung and retrocardiac areas were 13%-36% lower than published val ues for CaWO4. In the mediastinal area, there was no significant diffe rence between the CaWO4 and rare-earth phosphors. The use of grids red uces the scatter fraction in all three regions by 40%-75%. With and wi thout a grid, radiographic contrast with Gd2O2S was measured utilizing simulated lesions placed above the lung and retrocardiac areas.