Separation of mixtures containing hypophosphite, phosphite, orthophosp
hate, chloride, nitrate and sulphate ions has been achieved using a Vy
dac 300 IC 405 column, refractive index detection and eluents which we
re 2mM in p-hydroxy benzoic acid or 4mM in p-amino benzoic acid (adjus
ted to pH 5.5). Calibration curves were linear over the concentration
range 3 to 30 mg P/l.; and the precision (RSD for peak heights in mixt
ures) was < +/-10%. Using p-hydroxy benzoic acid, detection of eluted
peaks by indirect UV absorption was equally satisfactory. In an initia
l survey, the elution behaviour of nine oxyanions of phosphorus (simpl
e ions and condensed species) was examined, using eluents containing d
ifferent organic carboxylic acids (six aliphatic and nine aromatic), p
resent at levels of 1 to 5mM; with the pH adjusted to values between 3
and 8.6. No suitable conditions for separating mixtures containing co
ndensed phosphates were identified.