APPLICATION OF STANDARDLESS ANALYSIS IN GRAPHITE-FURNACE ATOMIC-ABSORPTION SPECTROMETRY - DETERMINATION OF CHROMIUM

Authors
Citation
Ys. Zheng et Xg. Su, APPLICATION OF STANDARDLESS ANALYSIS IN GRAPHITE-FURNACE ATOMIC-ABSORPTION SPECTROMETRY - DETERMINATION OF CHROMIUM, Talanta, 40(3), 1993, pp. 347-350
Citations number
14
Journal title
Talanta
ISSN journal
00399140 → ACNP
Volume
40
Issue
3
Year of publication
1993
Pages
347 - 350
Database
ISI
SICI code
0039-9140(1993)40:3<347:AOSAIG>2.0.ZU;2-2
Abstract
Influences of atomization temperatures on the characteristic mass and the atomic absorption coefficient of chromium were studied. The experi mental results show that the values of characteristic mass have appear ed to be stable to better than 10% when the analyte is atomized in the range of 2500-2800-degrees. The standardless analysis was applied to the determination of chromium in standard sediment and geochemical ref erence samples and satisfactory results were obtained.