A flow injection titrimetric method for the assay of vitamin C using c
erium(IV) was developed. A 157 mul vitamin C solution was injected int
o a carrier stream of 7.48 x 10(-3) M cerium(IV) prepared in 0.10 M su
lfuric acid flowing at a rate of 5.0 ml/min. A 1.3 ml mixing chamber w
as employed for dispersion of the resultant solution which was propell
ed through a flow-through cell and detected using the spectrophotomete
r at a wavelength monitored at 410 nm. The method was applied to the d
etermination of vitamin C in pharmaceutical preparations and a high ac
curacy was attained when compared with the British Pharmacoepeia (BP)
method. Excellent precision indicated by a relative standard deviation
of less than 0.9% was always obtained. No interferences were encounte
red and as many as 36 samples per hour were reached for throughput.