Mr. Matthews et Pj. Parsons, A SIMPLE METHOD FOR THE DETERMINATION OF LEAD IN WINE USING ZEEMAN ELECTROTHERMAL ATOMIZATION ATOMIC-ABSORPTION SPECTROMETRY, Atomic spectroscopy, 14(2), 1993, pp. 41-46
A simple method for the determination of lead in wine by electrotherma
l atomization AAS is described. Lead is measured following a simple 1:
4 dilution of wine samples with a chemical modifier solution containin
g 0.25% HNO3, 0.25% NH4H2PO4, and 0.0125% Mg(NO3)2. Instrument calibra
tion is carried out using six Pb standard solutions that are matrix-ma
tched for ethanol content. The peak appearance times for Pb in standar
ds and wine samples are similar, thus permitting calibration with matr
ix-matched aqueous Pb standards, Using the optimized furnace condition
s, the characteristic mass (m0) is 41.7 pg (expected mo = 33 pg). Over
all Precision is good (RSD less-than-or-equal-to 5.0%) and the minimum
detection limit (3sigma) for Pb in wine is 6.2 mug/L. The accuracy of
the method is assessed through participation in an interlaboratory co
mparison of methods for the determination of Pb in wine organized join
tly by the European Community's Bureau of Community Reference (BCR) an
d the U.S. Food and Drug Administration (FDA). The certified values ar
e within the confidence limits reported with results obtained using th
is method. Results of analysis of several American and European wines
are reported. The concentrations of Pb in wines tested were relatively
low compared with the European recommended limit of 300 mug/L for Pb
in wine, although some New York State and European wines approached or
exceeded a proposed U.S. limit of 50 mug/L.