MALPOSITION OF EXTRACORPOREAL MEMBRANE-OXYGENATION CANNULAS IN PATIENTS WITH CONGENITAL DIAPHRAGMATIC-HERNIA

Citation
K. Raisbahrami et al., MALPOSITION OF EXTRACORPOREAL MEMBRANE-OXYGENATION CANNULAS IN PATIENTS WITH CONGENITAL DIAPHRAGMATIC-HERNIA, The Journal of pediatrics, 122(5), 1993, pp. 794-797
Citations number
12
Categorie Soggetti
Pediatrics
Journal title
ISSN journal
00223476
Volume
122
Issue
5
Year of publication
1993
Part
1
Pages
794 - 797
Database
ISI
SICI code
0022-3476(1993)122:5<794:MOEMCI>2.0.ZU;2-I
Abstract
We describe two infants with congenital diaphragmatic hernia who under went extracorporeal membrane oxygenation and in whom the venous cannul a was in the left atrium instead of the right. The routine radiograph of the chest failed to demonstrate the malposition. We recommend using the echocardiogram to confirm the position of the cannula or to guide the surgeon during the cannulation of patients with congenital diaphr agmatic hernia.