EVALUATION OF SENSITIVITY OF 10 DIAGNOSTIC ASSAYS FOR CHLAMYDIA-TRACHOMATIS BY USE OF A SIMPLE LABORATORY PROCEDURE

Citation
Bj. Thomas et al., EVALUATION OF SENSITIVITY OF 10 DIAGNOSTIC ASSAYS FOR CHLAMYDIA-TRACHOMATIS BY USE OF A SIMPLE LABORATORY PROCEDURE, Journal of Clinical Pathology, 46(5), 1993, pp. 408-410
Citations number
6
Categorie Soggetti
Pathology
ISSN journal
00219746
Volume
46
Issue
5
Year of publication
1993
Pages
408 - 410
Database
ISI
SICI code
0021-9746(1993)46:5<408:EOSO1D>2.0.ZU;2-Y
Abstract
Aims-To determine the sensitivity of commercially available diagnostic assays for Chlamydia trachomatis using a simple method. Methods-Nine commercial assays and an ''in-house'' polymerase chain reaction (PCR) were evaluated using serial dilutions of a laboratory grown H serovar- four of them using a laboratory grown E serovar. Seven of the assays w ere further tested using dilutions of several cervical samples known t o contain chlamydiae. Results-The most sensitive assays were the Micro Trak direct fluorescent antibody (DFA) test (Syva) and the PCR which d etected C trachomatis at a 10(-8) dilution of the H serovar, while the two least sensitive, Clearview (Unipath) and TestPack (Abbott), were positive only at 10(-4) and-3 dilutions, respectively. A range of enzy me immunoassays (EIAs) and a nucleic acid hybridisation test were of i ntermediate sensitivity. The results with serovar E were consistent wi th these. When clinical samples were examined, the DFA test detected C trachomatis in dilutions at least 10-fold greater than any other assa y. Conclusions-The range of sensitivity of diagnostic assays determine d by the laboratory dilution procedure is very wide. Sensitivity asses sed in this way, however, reflects the ability of the assays to detect C trachomatis in large scale clinical trials. The dilution procedure, which is simple to undertake, could therefore be applied by any labor atory before a new diagnostic method is considered for routine use.