EXIT SWITCH FOR RADIOTHERAPY TREATMENT ROOMS

Authors
Citation
G. Manson, EXIT SWITCH FOR RADIOTHERAPY TREATMENT ROOMS, Physics in medicine and biology, 38(5), 1993, pp. 639-641
Citations number
3
Categorie Soggetti
Radiology,Nuclear Medicine & Medical Imaging
ISSN journal
00319155
Volume
38
Issue
5
Year of publication
1993
Pages
639 - 641
Database
ISI
SICI code
0031-9155(1993)38:5<639:ESFRTR>2.0.ZU;2-M
Abstract
An improved circuit design for a radiotherapy treatment room exit alar m and interlock is described and compared with simpler circuits. The u nique self test and fail safe features are discussed. All component va lues are given for the unit which has been in successful operation for over one year.