HIGH-PERFORMANCE LIQUID-CHROMATOGRAPHIC DETERMINATION OF PERPHENAZINEIN PLASMA

Authors
Citation
Tk. Mandal et Ln. Ace, HIGH-PERFORMANCE LIQUID-CHROMATOGRAPHIC DETERMINATION OF PERPHENAZINEIN PLASMA, Journal of clinical pharmacy and therapeutics, 18(3), 1993, pp. 205-208
Citations number
9
Categorie Soggetti
Pharmacology & Pharmacy
ISSN journal
02694727
Volume
18
Issue
3
Year of publication
1993
Pages
205 - 208
Database
ISI
SICI code
0269-4727(1993)18:3<205:HLDOP>2.0.ZU;2-A
Abstract
A reversed-phase high-performance liquid chromatographic (HPLC) method for the analysis of perphenazine in rabbit plasma was developed. Amit riptyline HCl was used as the internal standard. The chromatography wa s performed using a Microsorb-CN column; the mobile phase consisted of 85:15 methanol/0.005 M ammonium acetate buffer; flow rate of 1.6 ml/m in; and ultraviolet detection at 254 nm. Retention times were 10.8 and 18.1 min for perphenazine and amitriptyline HCI, respectively. The re covery of perphenazine added to plasma in the concentration range of 0 .05-5 mug/ml was found to be greater than 80%. This procedure was used for the analysis of plasma samples collected over time following the oral administration of perphenazine tablets in rabbits.