J. Persliden et al., COMPARISON OF IMAGE QUALITY AND MEAN ABSORBED DOSE TO THE BREAST FOR 2 MAMMOGRAPHIC FILMS, Acta radiologica, 34(4), 1993, pp. 351-355
A film intended for mammography (Kodak SO 155 MRH-1) was tested and co
mpared to one (Kodak SO 177 Ortho M) used earlier. Both films includin
g the cassette could resolve 20 1p/mm. For a processing time of 90 s t
he new film gave the same overall image quality and irradiation dose t
o the breast as the old system. If, instead, the processing time was i
ncreased to 150 s, a 43% reduction in kerma could be attained with the
new film. With a developing temperature of 36-degrees-C and 150 s pro
cessing time, the noise is clinically acceptable. A nonparametric test
showed no significant difference between the 2 films on the 0.01 leve
l. At an X-ray tube potential difference of 25 kV, the mean absorbed d
ose to a 4.5-cm-thick breast was reduced from 1.7 mGy with the old com
bination to 1.0 mGy with the new one. The measurements were made with
a moving grid.