EVALUATION OF MAMMOGRAPHIC VIEWBOX LUMINANCE, ILLUMINANCE, AND COLOR

Citation
Ag. Haus et al., EVALUATION OF MAMMOGRAPHIC VIEWBOX LUMINANCE, ILLUMINANCE, AND COLOR, Medical physics, 20(3), 1993, pp. 819-821
Citations number
9
Categorie Soggetti
Radiology,Nuclear Medicine & Medical Imaging
Journal title
ISSN journal
00942405
Volume
20
Issue
3
Year of publication
1993
Pages
819 - 821
Database
ISI
SICI code
0094-2405(1993)20:3<819:EOMVLI>2.0.ZU;2-B
Abstract
Twenty-three viewboxes were evaluated in six mammography facilities. L uminance and illuminance measurements were made with a recently calibr ated photometer. Color temperatures were measured with a Minolta color meter. The average luminance for the 23 viewboxes was 2920 nit (lumen /Sr/m2), the lowest value was 1610 nit, and the highest value was 3630 nit. The average illilminance was 40 lux (lumen/m2), with 6 lux as th e lowest value, and 97 lux as the highest. The average color temperatu re was 8400 K with the lowest value of 4900 K, and the highest of 10 9 00 K.