A NONPENETRATING FETAL SCALP ELECTRODE

Citation
Gj. Hofmeyr et al., A NONPENETRATING FETAL SCALP ELECTRODE, British journal of obstetrics and gynaecology, 100(7), 1993, pp. 649-652
Citations number
9
Categorie Soggetti
Obsetric & Gynecology
ISSN journal
03065456
Volume
100
Issue
7
Year of publication
1993
Pages
649 - 652
Database
ISI
SICI code
0306-5456(1993)100:7<649:ANFSE>2.0.ZU;2-1
Abstract
Objective To develop a nonpenetrating scalp electrode for intrapartum fetal monitoring. Design Preliminary observational evaluation of the d evice. Setting An urban academic hospital in Johannesburg, South Afric a. Subjects Fifteen women in labour. Intervention Application of the e lectrode for intrapartum monitoring. Outcome measures Quality of traci ngs, duration of application, disconnection of the electrode, marking of the scalp. Results Application was successful in 13 of 15 women, an d high quality tracings obtained in 12. Conclusions The design of scal p electrode tested is effective and produces high quality tracings in most cases.