IMPROVED ACOUSTIC-PLATE-MODE BIOSENSOR

Citation
Jc. Andle et al., IMPROVED ACOUSTIC-PLATE-MODE BIOSENSOR, Sensors and actuators. B, Chemical, 13(1-3), 1993, pp. 437-442
Citations number
5
Categorie Soggetti
Engineering, Eletrical & Electronic","Instument & Instrumentation
ISSN journal
09254005
Volume
13
Issue
1-3
Year of publication
1993
Pages
437 - 442
Database
ISI
SICI code
0925-4005(1993)13:1-3<437:IAB>2.0.ZU;2-B
Abstract
The design of an improved acoustic-plate-mode (APM) biosensor is prese nted. The biosensing element consists of a pair of APM delay lines fab ricated on an ZX lithium niobate (LiNbO3) substrate. Each delay line u tilizes a single APM and exhibits adjacent-mode rejection superior to previous LiNbO3 designs, without sacrificing the high sensitivity to e lectrical effects. The biosensing element is evaluated as an immunosen sor using the vector voltmeter (VVM) measurement system. The results s how that the improved biosensor is capable of selectively detecting an tibody concentrations of 20 ng per ml.